Home News About Us Contact Us

Industrial / Bio-Pharm:

LabPure 150

LabPure 250 EV

LabPure 500 EV

LabPure 500

LabPure 1000

R.O. Custom Design

 

Back to Water Systems

LabPure 1000

Design Includes

Automatic low pressure shut-off sensor for pump protection and blocked pre-filter indication.
Re-pressurization pump for water pressure control to R.O. membranes;
Automatic feed water shut off when reservoir is full.
Automatic flush cycle 2 minutes every hour to regenerate the system.
Preloaded with:
a) 25 - 1m Spun Wound Gradient Density Pre-Treatment Filter;
b) 1m, Extruded Carbon Block Pre-Filter.
High flow/High rejection 1000 GPD R.O. membrane;
45 gallon all natural polypropylene reservoir with level limit float switch, free standing.
A stand with industrial casters.
Size: 30"L × 26"W × 72"H.

Click to enlarge
Item Catalogue Number Price, CA$
LabPure 1000 AS-LABPURE1000 Request

Benefits

1000 US gallon per day (GPD) capacity (2700mL/min at 110 PSI and 7-8C).
Operates on standard 115V/15Amp power socket.
Ideal to replace older stills and R.O. systems.
Easy installation, low cost of maintenance.
Sound proof for quiet operation.
C.S.A. certified.
One-year warranty – parts and labour.
Made in Canada.

Water Requirements

Cold city water ½" line with 40-80 PSI feed pressure terminated with a ½" NPTF ball valve connection.
Water consumption is 3 US GPM.
Drain should be capable of handling 4 GPM.

Standard Expendables

Item Approximate Service Life Quantity Catalogue Number Price, CA$
Pre-Treatment Filter (1m, Spin Wound, Gradient Density) - 1 pack PF2501-20 Request
Pre-Filter (1m, Extruded Carbon Block) - 1 pack FF1C4-20 Request
Post-Filter (Mixed Bed, Nuclear Grade 20" D.I. - 1 pack DI2-20 Request
Final Filter (0.22m, Sterile, Pharmaceutical Grade) - 1 pack ZEMSB020NNPS3 Request
(2) High Flow/Rejection R.O. Membrane - 1 pack AG4021 Request

Accessorises

Item Quantity Catalogue Number Price, CA$
190 Gallon Self-Standing Reservoir Tank
7" Vent Port
Low-Level Flow Switch Control
1" Bulkhead Feed Water Connection
1" Bulkhead Overflow
Valves, Fittings and Tubing
Overall Size: 31" Diameter × 80" Height
1 SCTC190 Request
Re-Pressurization Pump, 20 GPM 1 SCFS1012 Request
VFD Re-Pressurization Pump, 1-5 GPM 1 SCVFD1-3 Request
Re-Circulation Pump, 5 GPM, continuous duty. Other pumps are available. 1 SCGFS1005 Request

Phone (403) 287-2345 Toll-free 1-(800)-565-2345 Fax (403) 287-1355 E-mail: sincaninc@telus.net